Contraindications, Warnings, and Cautions
CONTRAINDICATIONS:
Do not use the 950 in an MRI environment.
Do not operate the 950 in an explosive atmosphere.
Do not use the 950 on infant or neonatal patients.
WARNINGS:
Batteries may leak or explode if used or disposed of improperly.
CAUTIONS:
Federal Law (USA) restricts this device to sale by or on the order
of a physician.
Read this entire manual carefully before using the Respironics 950
Pulse Oximeter.
The 950 determines the percentage of arterial oxygen saturation of
functional hemoglobin. Significant levels of dysfunctional
hemoglobin such as carboxyhemoglobin or methemoglobin may
affect the accuracy of the measurement.
Cardiogreen and other intravascular dyes, depending on the
concentration, may affect SpO
2
accuracy.
The 950 must be able to measure the pulse properly to obtain
accurate SpO
2
measurement. Verify that nothing is hindering the
pulse measurement before relying on the SpO
2
measurement.
Some situations that may contribute to erroneous pulse
measurement are:
inadequate perfusion,
excessive patient movement,
pulse oximeter and blood pressure measurement on the
same limb, etc.
The 950 may interpret motion artifact of sufficient amplitude and
regularity as good perfusion (green).
Do not use the 950 in situations where alarms are required. The
950 has no audible alarms.
The 950 is intended for spot-checking or continuous monitoring by
an attending health care professional. Because the 950 has no
audible alarms, international labeling requirements (ISO 9919)
dictate it be labeled, “Not for continuous monitoring”.